excellence in High Voltage photo of LS Series unit
AppliedKilovolts
LS series datasheet
LS001, LS2.5, LS005, LS010, LS015, LS020, LS030 & LS050
PRECISION HIGH VOLTAGE MODULES
   1kV, 2.5kV, 5kV, 10kV, 15kV, 20kV, 30kV, 50kV
   Highest Stability, Lowest Ripple, Lowest Drift
   Externally programmable or Internal control (option)
   Short circuit and flashover proof
   24 hour burn in
   Low radiated magnetic field
   Positive, negative & remote reversible versions
applications (1 Watt max):
   mass spectrometers
   electron microscopes
   nuclear instruments
   photomultiplier tubes
   microchannel plates
The LS series sets a new standard for high voltage precision and stability.   By optimising the "Low Stress" technologies used in our leading 10 Watt HP Series for 1 Watt applications, the new LS series has less th an 1ppm ripple, 10ppm/°C temperature coefficient, and 50ppm drift/hr throughout the range.   Each of the units includes a differential control input.   To take full advantage of the performance available from the LS range, they are initially only being made available to our OEM customers as application specific, custom optimised, modules.   All units are short circuit proof and use proven techniques to drive a high frequency oscillator and ferrite high voltage step-up transformer.   These advanced power supplies build on Applied Kilovolts' legendary reputation for reliability, being constructed from conservatively rated components with their reliability further enhanced by information gained over many years of field operation.   Available as positive and negative units up to 50kV, and electrically reversible at voltages up to 30kV.   Please see the HPZ range for zero crossing reversible (bi-polar) units.

pin connections & safety instructions

LS series (unipolar) electrical specification
unit type polarity output ripple at full load dimensions (mm)
LS001PIP010 positive 10V to 1kV   @1mA 2mV   peak to peak 156 x 127 x 64
LS001NIP010 negative
LS2.5PIP010 positive 10V to 2.5kV   @400µA 3mV   peak to peak 156 x 127 x 64
LS2.5NIP010 negative
LS005PIP010 positive 10V to 5kV   @200µA 5mV   peak to peak 156 x 127 x 64
LS005NIP010 negative
LS010PIP010 positive 20V to 10kV   @100µA 10mV   peak to peak 156 x 127 x 64
LS010NIP010 negative
LS015PIP010 positive 30V to 15kV   @66.7µA 15mV   peak to peak 207 x 148 x 74
LS015NIP010 negative
LS020PIP010 positive 50V to 20kV   @50µA 20mV   peak to peak 207 x 148 x 74
LS020NIP010 negative
LS030PIP010 positive 100V to 30kV   @33.3µA 30mV   peak to peak 207 x 148 x 74
LS030NIP010 negative
LS050PIP010 positive 200V to 50kV   @20µA 50mV   peak to peak TBD
LS050NIP010 negative

LSR series (reversible) electrical specification
unit type polarity output ripple at full load dimensions (mm)
LS001RIP010 reversible +/-10V to +/-1kV   @1mA 2mV   peak to peak TBD
LS2.5RIP010 reversible +/-10V to +/-2.5kV   @400µA 3mV   peak to peak TBD
LS005RIP010 reversible +/-10V to 5kV   @200µA 5mV   peak to peak TBD
LS010RIP010 reversible +/-20V to +/-10kV   @100µA 10mV   peak to peak TBD
LS020RIP010 reversible +/-50V to +/-20kV   @50µA 20mV   peak to peak TBD
LS030RIP010 reversible +/-100V to +/-30kV   @33.3µA 3mV   peak to peak TBD

ELECTRICAL SPECIFICATION
INPUT VOLTAGE: - +24 volt d.c. +/-10% at less than 150mA.   Negative  input terminal common to HV ground return.
OUTPUT VOLTAGE: - see separate table.
LINE REGULATION: - Less than 2ppm of maximum voltage for input changes of 1 volt.
LOAD REGULATION: - Less than 10ppm of maximum voltage for load changes from 10% to 100% full load. (Measured at maximum voltage).
OVERLOAD PROTECTION: - Flashover and short circuit proof.   Re-entrant current limit.
TEMPERATURE CO-EFFICIENT: -    10ppm / °C ( 20°C ~ 50°C )
DRIFT: - < 50ppm / 1 hour (< 0.005%)  
< 100ppm / 8 hours (< 0.01%)  
measured at constant input voltage, load current & ambient temperature after 0.5 hour warm up.
CONTROL: 0 to +10V gives zero to maximum output, tolerance +/-2%
input impedance > 100kohm
  • INTERNAL or EXTERNAL potentiometer --- OPTION
VOLTAGE READOUT: - 0V ~ 10V represents zero ~ maximum output, tolerance +/-2%   ( source resistance 10kohm ).
CURRENT MONITOR: - Optional on all units.   Precision current monitor: 0 ~ 10V represents zero to 100% o/p current, tolerance +/-2%, offset +/-0.1% of FS.
OPERATING TEMPERATURE: - 0°C to +50°C
STORAGE TEMPERATURE: - -35°C to +85°C
R.F.I.: - steel case for low radiated magnetic field.
MECHANICAL SPECIFICATION
FIXINGS: blind fasners   front and rear
OUTPUT: by 1 meter screened cable, or optional integrated high voltage connector
N.B.   reducing the cable length may increase the ripple voltage.
INPUT CONNECTOR: 15 way D-type


AppliedKilovolts Applied Kilovolts Ltd.     http://www.appliedkilovolts.com/
  38 - 40 Eastbrook Road, Portslade, Sussex,
  BN41 1PB, United Kingdom
  Phone: +44 (0) 1273   439 440        Fax: +44 (0) 1273   439 449
Local distributor in Japan
APACE Science
APACE Science, Inc.     http://www.apace-science.com/
  Mitomo-Bldg.#2 - 6F, Ebisu-nishi 2-Chome,
  2 - 8, Shibuya-Ku, Tokyo 150-0021
  Phone: +81 (0)3 - 3463 - 3204       Fax: +81 (0)3 - 3463 - 3205  

| page top | standard modules | pin out | current monitor | APACE Science Home |